Comparison between the role of transabdominal ultrasound versus transvaginal ultrasound in evaluation of placental invasion in cases of placenta previa anterior wall with previous uterine scar

Document Type : Original Article

Authors

1 Department of Obstetrics & Gynecology, Faculty of Medicine, Al-Azhar University, Egypt

2 Department of Obstetrics & Gynecology, Faculty of Medicine, Al-Azhar University, Egypt

Abstract

Background: Maternal mortality from placenta accrete is estimated to be 6-7 % regardless of the type of the operation.
Aim Of The Study :To evaluate the influence of transabdominal versus transvaginal ultrasound in assessing placental invasion in situations of placenta previa anterior wall with a prior uterine scar using the unified ultrasonographic characteristics proposed by "EW-AIP," as well as to assess the sensitivity and accuracy of each characteristic by evaluatingthem to the pregnancy''s end results.
Patients and Methods: A sum of 100 pregnant female with persisting placenta previa (beyond 28 weeks of pregnancy) were included in this research. Transabdominal and transvaginal ultrasonography were conducted by two separate physicians who were unaware of each other''s findings. TAS and TVS applied and analyzed unified descriptors on the placenta in order to determine its precise location.
Results: At the time of Cesarean delivery, 86 individuals had unusually invasive placentas and variations, which were subsequently validated by histological study. The reliability of diagnosis of the retroplacental clear zoneloss was 76 percent by TVS and 54 percent by TAS, while that of aberrant placental lacunae was 92 percent by TVS and 88 percent by TAS. The reliability of detecting myometrial thinning was 66 percent by TVS and 72 percent by TAS.
Conclusion: Transabdominal and transvaginal ultrasound have been proven to be complimentary to each other, with transvaginal ultrasound having the upper hand. The safety of TVS has also been established, and the unified descriptors have been proven to be dependable in correct diagnosis.

Keywords


INTRODUCTION

By enabling us to pick the ideal time and location of delivery, prenatal identification of Morbidly Adherent Placenta (MAP) and its variations may help decrease fetus/maternal sickness and death. Early diagnosis of placental disease allows for multidisciplinary surgical intervention, newborn critical care, preventive hypogastric artery balloon closure, embolization of the uterus, and an appropriate amount of blood units in the operation theatre.1

The diagnosis of sonographic criteria for placenta accrete has been examined by many authors. Always evaluate the danger of undergoing a needless hysterectomy (false positive) or subsequent hemorrhage after attempted placental removal (false

 

negative). The use of diagnostic criteria cannot be justified merely on the basis of sensitivity and specificity; PPV and NPV must be assessed in order to plan appropriate patient treatment and information.2

A study of the literature from the previous decade suggests that placenta accreta is becoming more common, owing to more common CSs. A prior uterine scar is nearly often the location of aberrant placental invasion. Greater maternal age, myometrial injury after a myomectomy with endometrial entry, extensive curettage with subsequent Ashermandisorder, and submucosal myoma are all risk factors that are less significantly linked to MAP.1

Women who have had a prior CS with a placenta previa atop the former uterine scars are at the biggest danger of an unusually invasive placenta.3

A variety of distinct ultrasonography criteria, some qualitative and others quantitative, are used to diagnose a morbidly adhered placenta. Several investigations have looked at the accuracy of several ultrasonography markers for placental attachment disorders (PAD). However, the efficacy of these markers varies a lot amongst investigations that use the same indicators. Limits on research population, retrospective methodology, and research requirements variation, as well as the eventual determination of AIP, have all been blamed for the differences.4

Moreover, like with other subjective diagnostic approaches, what defines a marker will influence whether or not each symptom is logged, according to the operator. This is especially crucial for practitioners who may not have much expertise with placental ultrasound or AIP diagnosis. Furthermore, there is no documented agreement on the characterization of the ultrasonography indicators that are typically used to diagnose AIP. Many indications have been recognized with several names, and in other instances, the same title has been used to many discoveries.5

The European Working Group on Unnaturally Invasive Placenta (EW-AIP) is a non-profit association with 29 members from 11 European countries, including obstetricians, gynecologists, pathologists, anesthesiologists, and basic science researchers. The group's mission is to enhance AIP diagnosis and treatment while also promoting research and understanding. The EW-AIP provides standardized definitions of the AIP imaging descriptors to enhance the comparability of future research, expand diagnostic capabilities, and promote international cooperation. All 23 papers in a recent systematic review of the prenatal sonographic detection of AIP were evaluated to establish these standardized criteria. 5

Six descriptions for 2D grayscale ultrasound, four for 2D color Doppler, and one for 3D power Doppler were created from the various wordings. The significance of describing each indication clearly was put on it, regardless of views about its predictive usefulness. 5

The aim of this research is to evaluate the influence of transabdominal versus transvaginal ultrasound in assessing placental invasion in situations of placenta previa anterior wall with a prior uterine scar using the unified ultrasonographic characteristics proposed by "EW-AIP," as well as to assess the sensitivity and accuracy of each characteristic by evaluating them to the pregnancy's end results..

PATIENTS AND METHODS

This was a prospective Observational Cohort research with 100 pregnant women with persisting placenta previa (beyond 28 weeks' pregnancy) who attended Al-Hussien Hospital between April 2021 and the conclusion of the research adhering to the below inclusion and exclusion criteria:

Inclusion Criteria:Pregnant women identified with persisting placenta previa anterior wall following 28 weeks of pregnancy, and a histories of prior cesarean deliveryand/or additional uterine procedures (age group 18-45 years).

Exclusion Criteria:Unscarred uterus and placenta previa posterior wall

All the patients were submitted to the following steps:

Complete history, including:Personal history (name, age, file number).Obstetric history (number of C.S., abortion, prior gestation placenta previa, ectopic gestation history, medical disorder with pregnancy and number of living children).Present history (complain, gestational age, medical disorder in present pregnancy and history of ante partum hemorrhage), and past history (postpartum sepsis, postpartum hemorrhage & chronic diseases).

Two skilled operators performed transabdominal and transvaginal imaging utilizing all diagnostic procedures (gray-scale, color Doppler), followed by an offline analysis of the collected pictures and volumes.

Intraoperative information, such as placental site, spontaneous separation, and placental invasion of the bladder and other organs. Administration during pregnancy. Injury to the bladder, ureter, or bowel. Blood transfusions and blood loss to prevent a hysterectomy and the requirement for critical care unit hospitalization, conservative treatment is used.

Histopathological examination in cases of hysterectomy.

Sample size justification:

According to Sedek et al.6, the overall accuracy of TVS identification was around 97.1 percent when using the EPI Info 7 software for sample size calculating. With a margin of error of 5% and a confidence level of 95 percent, a sample size of at least 100 women was required.

Statistical Analysis:

The statistical program for social sciences, version 23.0, was utilized to analyze the data (SPSS Inc., Chicago, Illinois, USA). The quantitative data was given in the form of a median, standard deviation, and range. The median and standard deviation were used to convey quantitative data (SD). Frequency and percentage were utilized to convey qualitative data. When comparing two averages, an independent-samples t-test of relevance was employed. To compare percentages between two qualitative factors, the Chi-square (X2) test of relevance was applied. For data correlation, Pearson's correlation coefficient (r) test was employed.Evaluation of Diagnostic Performance for histopathology and Transabdominal ultrasound & Transvaginal ultrasound. Sensitivity = (true +ve)/ [(true +ve) + (false –ve)]. Specificity = (true –ve) / [(true –ve) + (false +ve)]. PPV = (true +ve) / [(true +ve) + (false +ve)]. NPV = (true –ve)/ [(true –ve) + (false –ve)]. Accuracy = (TP+TN)/[TP+FP+TN+FN]. McNemar's test was used to examine the relationship between two (paired) qualitative variables. The tolerable margin of error was set at 5%, while the confidence interval was set at 95%.A P value of less than 0.05 was deemed substantial.

 

RESULTS

Abnormal lacunea TAS

Histopathology Evaluation1

Total

Abnormal

No abnormal

Yes

Count

80

2

82

% of TAS

97.6%

2.4%

100.0%

% of Histopathology

93.0%

14.3%

82.0%

% of Total

80.0%

2.0%

82.0%

No

Count

6

12

18

% of TAS

33.3%

66.7%

100.0%

% of Histopathology

7.0%

85.7%

18.0%

% of Total

6.0%

12.0%

18.0%

Total

Count

86

14

100

% of TAS

86.0%

14.0%

100.0%

% of Histopathology

100.0%

100.0%

100.0%

% of Total

86.0%

14.0%

100.0%

Table 1: Comparison between Histopathology Evaluation1 and TAS according abnormal lacunae.

Table 1 shows statistically substantial diagnostic performance of Abnormal lacunae TAS it was sensitivity 93%, specificity 85.7%, positive predictive value 14.3% and negative predictive value 7% and accuracy 92%.

TAS

TVS

Total

Correct

Incorrect

 Correct

Count

52

2

54

% within TAS

96.3%

3.7%

100.0%

% within TVS

68.4%

8.3%

54.0%

Incorrect

Count

24

22

46

% within TAS

52.2%

47.8%

100.0%

% within TVS

31.6%

91.7%

46.0%

Total

Count

76

24

100

% within TAS

76.0%

24.0%

100.0%

% within TVS

100.0%

100.0%

100.0%

McNemar Test: p-value

<0.001

Table 2: A comparison of TAS and TVS accuracyaccording to loss clear zone.

Table 2 reveal highly statistically substantialvariation between TAS and TVS according to loss clear zone, with p-value (<0.001).

TAS

TVS

Total

Correct

Incorrect

Correct

Count

86

6

92

% within TAS

93.5%

6.5%

100.0%

% within TVS

97.7%

50.0%

92.0%

Incorrect

Count

2

6

8

% within TAS

25.0%

75.0%

100.0%

% within TVS

2.3%

50.0%

8.0%

Total

Count

88

12

100

% within TAS

88.0%

12.0%

100.0%

% within TVS

100.0%

100.0%

100.0%

McNemar Test: p-value

0.289

Table 3: A comparison of TAS and TVS accuracyaccording to abnormal lacunae.

Table 3 shows no statistically substantial variation of TAS and TVS regarding abnormal lacunae, with p-value (>0.05 non-substantial).

TAS

TVS

Total

Correct

Incorrect

Correct

Count

58

8

66

% within TAS

87.9%

12.1%

100.0%

% within TVS

80.6%

28.6%

66.0%

Incorrect

Count

14

20

34

% within TAS

41.2%

58.8%

100.0%

% within TVS

19.4%

71.4%

34.0%

Total

Count

72

28

100

% within TAS

72.0%

28.0%

100.0%

% within TVS

100.0%

100.0%

100.0%

McNemar Test: p-value

0.286

Table 4: A comparison of TAS and TVS accuracyaccording to myometrial thining.

Table 4 shows no statistically substantial variation between TAS and TVS regarding myometrial thining, with p-value (>0.05 non-substantial).

TAS

TVS

Total

Correct

Incorrect

Correct

Count

82

2

84

% within TAS

97.6%

2.4%

100.0%

% within TVS

93.2%

16.7%

84.0%

Incorrect

Count

6

10

16

% within TAS

37.5%

62.5%

100.0%

% within TVS

6.8%

83.3%

16.0%

Total

Count

88

12

100

% within TAS

88.0%

12.0%

100.0%

% within TVS

100.0%

100.0%

100.0%

McNemar Test: p-value

0.289

Table 5: Comparison of the accuracy of TAS and TVS regarding uterovesical hypervascularity.

Table 5 shows no statistically substantial variation between TAS and TVS regarding uterovesical hypervascularity, with p-value (>0.05 non-substantial).

TAS

TVS

Total

Correct

Incorrect

Correct

Count

26

4

30

% within TAS

86.7%

13.3%

100.0%

% within TVS

76.5%

6.1%

30.0%

Incorrect

Count

8

62

70

% within TAS

11.4%

88.6%

100.0%

% within TVS

23.5%

93.9%

70.0%

Total

Count

34

66

100

% within TAS

34.0%

66.0%

100.0%

% within TVS

100.0%

100.0%

100.0%

McNemar Test: p-value

0.388

Table 6: Comparison of the accuracy of TAS and TVS according to subplacental hypervascularity.

Table 6 shows no statistically substantial variation between TAS and TVS regarding subplacental hypervascularity, with p-value (>0.05 non-substantial).

TAS

TVS

Total

Correct

Incorrect

Correct

Count

70

4

74

% within TAS

94.6%

5.4%

100.0%

% within TVS

92.1%

16.7%

74.0%

Incorrect

Count

6

20

26

% within TAS

23.1%

76.9%

100.0%

% within TVS

7.9%

83.3%

26.0%

Total

Count

76

24

100

% within TAS

76.0%

24.0%

100.0%

% within TVS

100.0%

100.0%

100.0%

McNemar Test: p-value

0.754

Table 7: Comparison of the accuracy of TAS and TVS regarding bridge vessels.

Table 7 shows no statistically substantial variation between TAS and TVS regarding bridge vessels, with p-value (>0.05 non-substantial).

TAS

TVS

Total

Correct

Incorrect

Correct

Count

40

12

52

% within TAS

76.9%

23.1%

100.0%

% within TVS

76.9%

25.0%

52.0%

Incorrect

Count

12

36

48

% within TAS

25.0%

75.0%

100.0%

% within TVS

23.1%

75.0%

48.0%

Total

Count

52

48

100

% within TAS

52.0%

48.0%

100.0%

% within TVS

100.0%

100.0%

100.0%

McNemar Test: p-value

1.000

Table 8: Comparison of the accuracy of TAS and TVS regarding placental lacunea feeder vessels.

Table 8 shows no statistically substantial variation of TAS and TVS regarding placental lacunae feeder vessels, with p-value (>0.05 non-substantial).

 

Transabdominal ultrasound

Transvaginal ultrasound

Sensitivity

94.2%

98.1%

Specificity

85.0%

93.8%

Positive predictive value

92.0%

97.0%

Negative  predictive value

65.2%

89.1%

Accuracy

91.5%

97.5%

Validity (out of 86 patients)

Detected 74

Detected 84

Table 9: Comparison of the accuracy of TAS and TVS according to overall finding.

Table 9 shows statistically significant diagnostic performance of overall, as for the transabdominal ultrasound it was sensitivity 94.2%, specificity 85%, positive predictive value 92% and negative predictive value 65.2% and accuracy 91.5%, while transvaginal ultrasound it was sensitivity 98.1%, specificity 93.8%, positive predictive value 97% and negative predictive value 89.1% and accuracy 97.5%, this indicates that the transvaginal ultrasound the most diagnosis for placental invasion.

 

DISCUSSION

The results showed that regarding the presence of abnormal placental lacunae, they showed Sensitivity 93.0%, specificity 66.7%, PPV 97.6%, NPV 85.7%, accuracy 90% by transabdominal ulrasonography (TAS) and Sensitivity 88.4%, specificity 85.7%, PPV 97.4%, NPV 85.7%, accuracy 88.0% by transvaginal ultrasonography (TVS).

Lacunae's PPV varies more across authors than other indicators; they've been described as sensitive and specific in some investigations but not in others.

This result corresponds with Comstock et al.3, who found them to be 93 percent sensitive in women at 20 weeks and beyond with a 93 percent PPV, whereas Cali et al.1 discovered the existence of aberrant lacune exhibited sensitivity 73.0 percent and specificity 86.7 percent. The total pooled sensitivity and specificity from 13 investigations of lacunar spaces detecting MAP were 77 percent and 95 percent, respectively, in a recent systematic review, with a total diagnostic accuracy of 88 percent.4

Regarding loss of the retroplacental clear zone

This research revealedthat TAS had a sensitivity of 51.2 percent, specificity of 71.4 percent, PPV 91.7 percent, NPV 19.2 percent, and accuracy of 54.0 percent, but TVS had a sensitivity of 74.4 percent, specificity of 85.7 percent, PPV 97.0 percent, NPV 35.3 percent, and accuracy of 76.0 percent. Unlike Cali et al.1revealed that Sensitivity was 90%, specificity was 80%, PPV was 57%, and NPV was 97% when the retropalcental clear zone was lost.However, Finberg and Williams7stated that the most of false positive findings are caused by the disappearance of the retroplacental clear zone, and the criteria should not be employed alone to determine the diagnosis.

Wong et al.8The lack of the clear area was discovered in 37 (65%) of women without placenta accreta and 100 percent of those with it. As a result, it is perceptive but not specific. Because of its significant negative predictive value, the clear space seems to be used primarily to efficiently exclude placenta accreta (NPV).

This study revealed that interruption of the bladder wall had sensitivity 4.7 percent, specificity 100.0 percent, PPV 100.0 percent, NPV 14.6 percent, accuracy 18.0 percent by TAS, and sensitivity 9.3 percent, specificity 100 percent, PPV 100 percent, NPV 15.2 percent, accuracy 22.0 percent by TVS, which agrees with Comstock et al.3, where this result had Sensitivity 20 percent, PPV 75 percent, and Wong et al8, sensitivity 11%, specificity 100%. Unlike Cali et al.1, who found that these criteria had a Sensitivity of 70%, Specificity of 100%, PPV of 100%, and NPV of 100%.

The poor sensitivity observed by Comstock et al.3 and Wong et al.8 might be due to the fact that not all women received transvaginal ultrasonography under the very strict settings employed by Cali et al.1. The authors of that huge research first found that 300 ml in the bladder gave the greatest view of uterine–bladder contact, and then instilled that quantity into each woman's bladder.

Regarding the uterovesical hypervascualrity using color doppler flow

This study shows Sensitivity 86.0%, specificity 71.4%, PPV 94.9%, NPV 45.5%, accuracy 84.0% by TAS and Sensitivity 93.0%, specificity 57.1%, PPV 93.0%, NPV 57.1%, accuracy 88.0% by TVS the sensitivity of this descriptor significantly increased using TVS. Also, this agrees with Cali et al.1 where is revealed Sensitivity 95.0%, specificity 100%, PPV 100%, NPV 97%.

Regarding the occurrence of aberrant arteries connecting the placenta to the bladder wall

This study showed sensitivity 69.8 percent, specificity 100 percent, PPV 100 percent, NPV 35.0 percent, and accuracy 74.0 percent according to TAS, and sensitivity 71.2 percent, specificity 100 percent, PPV 100 percent, NPV 36.8 percent, and accuracy 76.0 percent according to TVS., which also confirms with Cali et al.1.

Although there have been no research showing the diagnostic performance of transabdominal vs. transvaginal ultrasound in the setting of presumed placental invasion, transvaginal ultrasound allows for a more thorough examination of the lower uterine segment and is the commonly prescribed quality of care.9

This research employed the criteria of the "EW-AIP" to examine the function of transabdominal ultrasonography vs. transvaginal ultrasound in assessing placental invasion in instances with placenta previa anterior wall with preexisting uterine scar.9

Throughout the period of this study, every patient who was enrolled has undergone both Transabdominal sonography and transvaginal sosngraphy and every one of the unified descriptors was assessed via both modalities, the accuracy of both modalities was calculated regarding their ability to evaluate each one of the unified descriptors as follows

The accuracy of TVS and TAS in detecting the disappearance of the retroplacental clear zone was 76 percent and 54 percent, respectively. TAS reported 92 percent of aberrant placental lacunae while TVS reported 88 percent. TAS had a detecting accuracy of 66% and TVS had a detecting accuracy of 72% for myometrial thinning. The accuracy of identification of uterovesical hypervascularity was 84 percent by TAS and 88 percent by TVS, according to the Doppler evaluation. TAS detected 76 percent of bridging vessels, whereas TVS detected 75 percent.

From this we conclude the accuracy of detection of the unified descriptors is quite close regarding TVS and TAS, however TVS was found to be more accurate in the exact placental localization putting in mind as well that The TVS was conducted by an extremely skilled operator, but the TAS was performed by less skilled operators. The total accuracy of identification for TAS was 91 percent, whereas the TVS was at 97.1 percent.

By enabling multidisciplinary counseling and delivery planning and scheduling, assigning a score in clinical practice may be beneficial in the prenatal identification of MAP and appear to be a crucial component in minimizing mother and fetus illness and death.

Other studies suggested different scoring systems based on different criteria that would eventually enhance the ability of ultrasonogropahy to predict abnormal placental invasion and thus enhance the offered approach.

CONCLUSION

This study suggests that both transabdominal and transvaginal ultrasonographic modalities are complementary to each other, putting in mind that TVS had a slightly higher overall accuracy and was done by a seasoned professional.

Also the unified descriptors suggested by the EW-AIP were found to be of dependable accuracy as well an important point to be considered is that all the patients who were enrolled in this study and were exposed to transvaginal ultrasound, none of them experienced any attack of bleeding during the procedure which proves the profound safety of this modality confirming many previous studies.

1. Calì G, Giambanco L, Puccio G, Forlani F. Morbidly adherent placenta: Evaluation of ultrasound diagnostic criteria and differentiation of placenta accreta from percreta. Ultrasound Obstet Gynecol. 41:406‑12. 2. Wong HS, Cheung YK, Williams E. Antenatal ultrasound assessment of placental/myometrial involvement in morbidly adherent placenta. Aust N Z J ObstetGynaecol. 2012; 52: 67–72 3. Comstock CH, Love JJ, Bronsteen RA, Lee W, Vettraino IM, et al. Sonographic detection of placenta accreta in the second and third trimesters of pregnancy. Am J Obstet Gynecol. 2011; 190: 1135–40. 4. D’Antonio F, Iacovella C, Bhide A. Prenatal identification of invasive placentation using ultrasound: systematic review and meta-analysis. Ultrasound Obstet Gynecol. 2013; 42: 509–17. 5. Collins SL, Ashcroft A, Braun T, Calda P, Langhoff‐Roos J, et al. Proposal for standardized ultrasound descriptors of abnormally invasive placenta (AIP). Ultrasound in Obstetrics & Gynecology. 2016; 47(3):271-5. 6. Sedek AA, Awad AM, Nafez MA. Comparison between the role of transabdominal ultrasound versus transvaginal ultrasound in evaluation of placental invasion in cases of placenta previa anterior wall with previous uterine scar. Al-Azhar Med J. 2019; 48(4): 335-50. 7. Finberg H, Williams J. Placenta accreta: prospective sonographic diagnosis in patients with placenta previa and prior cesarean section. J Ultrasound Med. 1992; 11:333-43. 8. Wong HS, Zuccollo J, Strand L, Tait J, Pringle KC. The use of ultrasoundin assessing the extent of myometrial involvement in partial placentaaccreta. Ultrasound Obstet Gynecol. 2007; 30:228-30. 9. Rac MW, Dashe JS, Wells CE, Moschos E, McIntire DD, et al. Ultrasound predictors of placental invasion: the Placenta Accreta Index. American Journal of Obstetrics and Gynecology. 2015; 212(3): 343-44.